| - The basic performance of the kit |
| Accuracy of the kit | |
| The basic performance of the Heparan Sulfate ELISA Kit was evaluated. |
| The difference between wells in a plate | |||||||||||||||||||||||||||||||||||||||||||||
| Three kinds of samples (urine) were measured according to the protocol using four wells per sample (in the same plate). | |||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||
| Difference between kits | |||||||||||||||||||||||||||||||||||||||
| Three kinds of samples (urine) were measured according to the protocol using three kits (from the same lot). | |||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||
| The difference between measurement dates. | |||||||||||||||||||||||||||||||||||||||||||||
| Three kinds of samples (urine) were measured on four different days according to the protocol using the kit (from the same lot). | |||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||
| The difference between lots. | |||||||||||||||||||||||||||||||||||||||
| Three kinds of samples (urine) were measured according to the protocol using the kit (three different lots). Measurement was carried out on the same day. | |||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||
| Conclusion | |
| The CV values are less than 10% in each of the above evaluations demonstrating good reproducibility of the kit. |